Reports until 17:21, Wednesday 09 April 2025
SQZ General
wenxuan - posted 17:21, Wednesday 09 April 2025 (11931)
VOPO crystal position characterization

Daniel, Wen

We did a characterization of the crystal position while optimizing temperature given the translation stage is working. The stage has a limit of [-6 mm, 4.5 mm] range. The edge of the crystal was measured to be [-2 mm, 2.7 mm] by sweeping the cavity and finding the crystal position where it lost resonance. The stage position readback is very stable and repeatable. 

We first attempted to perform the optimization while keeping the VOPO locked, in hope to see more sensitive NLG changes at a slightly different crystal position. The OPO loop could stay locked with the stage moving less than 0.01 mm. It turned out the NLG was not sensitive at the level of 0.01 mm, or roughly 0.1 C change in temperature (while keeping the other variable optimized). So we went back to rough characterization by sweeping the VOPO cavity. 

We started sweeping by sending in 5 Vpp with 2.5 V offset at 5 Hz. We were able to see the coresonance fairly quickly, and recorded the best temperature for all the positions of the crystal. We were able to see a very linear and repeatable function of the optimized temperature versus the crystal translation. The plot will be uploaded once Daniel's alog account is active. 

Today we went back to examine some of the outlier/missing data points from the last experiment. This time we increased the sweep range of the OPO PZT by decreasing the scanning frequency (getting away from the pole). We were able to pinpoint some of the missing data from last time. We do notice that there's a damaged spot on the crystal (labeled plot to be uploaded), which misaligns both green and red beams. We plan to do a finer NLG comparison at those good positions we found during the measurement.