[Begum, Haocun]
We assembled two additional homodyne boards for GRAVITES, labeled #5 and #6. #4 has been fitted with new diodes and requires further testing to confirm proper functionality. #5 is performing well and is currently in use. #6 appears to have lower gain, but I haven’t finished debugging yet.
The 1550 nm homodyne is now online for the rotating interferometer. We took dual-wavelength measurements with interfeormeter locked; Begum will follow up with detailed results.
[General] We saw abnormal 5 kHz peaks and their harmonics in the dark noise spectrum of the homodynes when measured with the SR785 in the EE shop, see here. I repeated the measurement using a different power supply, but the peaks remained. These oscillations do not appear when using the SR785 in the green lab, see here. The SR785 in the EE shop probably needs fixing.