General
sasha - posted 15:40, Thursday 03 April 2025 (11926)
Teraxion Laser Linearity Measurement + Intensity Noise
Continuing the testing of the Teraxion 1550 nm laser in the Yellow Lab, I recorded time series from an asymmetric Michelson with arm length difference of 17.5 cm, which is an increase from the previous 10 cm. I modulated the laser with a triangle wave (Vpp = 1V/2V, f = 1 kHz), and the response is shown in the plots (two for each voltage). I pressed on the table between measurements which changed the interference pattern, and that is the reason for big differences between measurements with the same modulation. 

For a more quantitative estimation of linearity, I also modulated the laser with a sine wave (Vpp = 1V/2V, f = 1 kHz) and will fit this data to calculate the effective modulation depth in higher harmonics. The results will be posted soon.

In addition, I measured the intensity noise, which appears to be mostly limited by the electronics noise of the photodetector (see the last plot)
Images attached to this report