The noise measured with the fiber interferometer mounted on the rotating arms on 2025-03-24, see plot. There is no discernible difference between the noise levels when the stage is powered versus not powered resting at 0 deg. Note the large acoustic noise coupling 100Hz to 10kHz, which may be concealing noise coupling in this region.
The attached plot also shows noise measurement done during rotation, which clearly shows acouastic coupling in the 100Hz-10kHz with jagged features. This experiment was repeated 4 times, with the system rotating 0 to 340 degrees and back, and the results seem quite consistent.
The second plot shows a second set of data taken on 2025-03-26 along with the data on 03-24. And the third plot shows the 03-26 data with SR785 dark noise. Low frequency side seems to be limited by this dark noise.
The last plot shows a comparison with previous data (March 13th and 14th data, log 11916), the acoustic noise coupling 100Hz to 10kHz doesn't look worse with the stage mounted configuration.
The next step is to fix the 1550 nm HD detector and get read-out.