dorotea - posted 05:10, Tuesday 25 March 2025 (11921)
Polarization control update

I've come to the conclusion that at least part of the issue I was having in getting the polarization control working in the past week was some kind of coupling going on in the pm-pm polarizers. I've switched these out for PBSs with an SM input and PM outputs. While using the polarizers, I observed changes in the polarization state measured by a thorlabs polarimeter directly from the output of the polarizer when manipulating the polarization state entering the polarizer, meaning they were not effectively polarizing the light. This caused issues at two points: first, when combining the measurement and control wavelengths at the start of the test setup, it meant that it was not possible to ensure both wavelengths started in the same polarization state-- meaning that the changes to the polarization caused by the controller would most likely have been totally different. In this scenario, even if the polarization of the 1542 light was completely stabilized, the polarization of the 1550 light would not be stable.  Second, since my control strategy is to minimize the light passing through a polarizer (ie, to drive the polarization of the light to be normal to the polarization axis of a polarizer), I was not getting a meaningful control signal.

The test setup is as follows: I have around 3mW of light coming in at either wavelength. Each passes through a set of polarization paddles, then the wavelengths are combined in a DWDM. The COM output of the DWDM passes through a SM-to-PM PBS. The fast axis output of the PBS enters a 1m SM fiber, passes through another set of paddles, then through the PCD-M02 polarization controller. The two wavelengths are then demuxed by a second DWDM. The 1542 output of the DWDM passes through a second SM-to-PM PBS. The rest of the setup is identical to the one I built in the fall at MIT: the slow axis output of the PBS is sent to a PD; the signal from the PD is split between a scope (for monitoring) and IN1 of a red pitaya. Note that the RP inputs don't like more than a volt when set to the low voltage mode. Always check the settings on the PDs and their TIA. IQ blocks are used to send a dither signal (currently set to about .1mV, 4kHz on OUT1 and 7kHz on OUT2). The demodulated signals are send to PID blocks, both of which currently use Kp = -1, Ki = -10. Finally, the output signals are sent through an amp circuit built by Tom (instead of the SR amp I used at MIT, but which performs the same operation of amplifying the RP output by a factor of ~2 and then adding 2.5V) and to the 1st and 2nd actuators on the PCD-M02 respectively. 

To monitor how well the loop works initially, I disconnected the 1550 light and replaced the 2nd DWDM with a 50/50 beamsplitter, of which one output is sent to the 2nd PBS and the other is sent to a thorlabs PAX1000 polarimeter so that I can directly monitor the polarization. I've tested the loops by slowly changing the polarization entering the PCD-M02 using the polarization paddles directly before it. I observe that, while the loops are running, the magnitude of the control signal coming from the PD is approximately minimized and the polarization is mostly stable as viewed on the polarimeter-- if I move the paddles a bit faster it does drift a little bit, but quickly returns to the same place-- the loops are definitely not optimized yet. Depending on the input polarization, the loops aren't able to get the signal all the way to the same minimum; hopefully I can fix this by optimizing the integer gains but I'm not sure that's the issue.

The video attached shows the two feedback signals (in red and green on the LHS of the screen), the polarization state of the 1542 light as measured by the PAX1000 polarimeter, and (at the bottom RHS of the screen) the DC voltage signal on the PD as I move the polarization around. If the polarization is moved slowly, the state stays pretty stable; if it's moved suddenly the state jerks a bit and then goes back to about the same state.

After taking the attached video I switched to a much longer (222m) SM fiber spool; this doesn't change the behavior of the loop much. However, after turning the 1550 back on and looking at the 1550 output of the second DWDM on the polarimeter, it seems that the loop is only able to constrain the polarization to within about 1/8th of the polarization sphere, not to the same polarization. The loops are definitely able to stabilize it somewhat, especially when the disturbances to the polarization are small and/or slow, but it's not perfect. This isn't too surprising; it's looking probable that we're going to need to figure out a scheme that uses the 1550 visibility at the output of the IFO to feed back on instead. 

I'll play with the integer values and take transfer functions of the loops in the near future but we should start seriously thinking about a new polarization control scheme for the moving iteration of the IFO; I'm hoping we can get away with this scheme for the stationary measurement.