Eric, Haocun, Wen
Following up on the 11830, we fixed the dichroic problems on the output of the squeezed vacuum. The problem is that both dichroics can individually give a factor of 50 (34 dB) of filtering of green while having low loss on red (< 1%). However, the filtering didn't really add up when put together. The problem is likely the AOI or polarization.
We used a power meter to measure the REFL of the 1st and 2nd in-chamber dichroics while adjusting the AOI of each optics. Given 5 mW of M1 REFL green, the 1st REFL was measured around 50 uW while the 2nd REFL was 10 uW. We then swapped the 1st and 2nd dichroic optics but still measured only a factor of 5 filtering on the 2nd dichoric. We swapped the order back given that the sequence didn't make any difference.
We then adjusted the 2nd dichroic optics by 90 deg in hope of improving the polarization. Then we saw 2 uW after 2nd, which was better than before. Haocun noticed that the power meter gave very inconsistent values with the internal ND filter on/off. We'll cross-calibrate it with another power meter.
We then switched to use a PD100A to measure the power. The red beam was blocked completely at the fiber input in addition to blocking the intracavity path. The 1st dichroic was measured with -27 mV with 8.2 V input and -71 mV dark offset, giving a filtering of around 45 dB (very good!). Then we measured -7 mV with 4.0 V input and -44 mV dark, giving 40 dB (a factor of 100) isolation. Now both dichroics are working as designed!
We still added an in-air 3rd dichoic to remove any of the DC pump light. We measured 77 mV before the dichroic and 65 mV transmitted. The filtered green was too small to measure, even with largest gain settting. We're pretty happy with the filtering, and we can now measure the NLG without extra pump landing on the TRANS PD.