RetroSense General
sasha - posted 19:09, Wednesday 19 March 2025 (11918)
Improved RetroSense Noise After Moving to the Yellow Lab + Connection Issue
I moved RetroSense v3 to the Yellow Lab and immediately saw an improvement in noise. The figure shows that we are now at or below the eBOSEM level at some frequencies and approaching shot noise. I also separately compared noise for two versions of the electronics setup, with and without demodulation, and it is very similar. With this, we are now performing similarly or better than v2 with having just the retroreflector, LED + QPD, and preamp, but no demodulation board. 

I measured this with the gain of 100 on SR560 to suppress the spectrum analyzer noise and making sure the QPD + LED are at the lens's focal length (12 mm). 

Having low photocurrent has been an issue in previous measurements, but today I noticed that loosening some of the screws in the optical head mount increases the photocurrent, and was able to increase it from <100 uA to 561 uA this way. This suggests a connection problem, and one of the LED connections was probably accidentally connected to ground. As a result, having a more mechanically robust optical head with standard connectors in place of hand-soldered wires would be the next improvement. 

dV/dx = 66.4 V/mm
V_sum = 14.3 V
I_sum = 561 uA
Shot noise = 8.22 pm/rtHz