Reports until 10:49, Monday 03 March 2025
GRAVITES General
begum - posted 10:49, Monday 03 March 2025 (11907)
Locking the interferometer on 1545nm path, with the prestabilization and pzt stretcher running

[Eric, Begum]

Eric had prepared two homodyne boards for the 1550 nm and 1545 nm paths. Last week, we wired them up and measured the dark noise. One of the boards had a reasonable noise level, see attached plot, whereas the other had elevated noise at low frequencies. So Eric replaced a resistor he suspected, and the noise became comparable to the other detector, see attached plot.
We locked the 1545 nm path of the interferometer, using the pzt strecher and feeding back to the VCO. One observation we had is the vaccum pump creating very fast fringe wrapping on the homodyne detector. (This is confirmed to be the cause since we turned it off for a short while and the fringe wrapping became much calmer.) The other observation is about the PZT strecther. It seems to affect the polarization state through the fiber, and hence the intensity of the fringes as well as the UGF. Eric mentioned he used epoxy to fix the the fiber on the pzt stretcher and we suspect this may be affecting the polarization while stretching. We will rewind the fiber on the strecher using double sided tape (like Haocun did in Vienna), and add more windings for larger range in length tuning. Ultimately we may need to set up active polarization control, but for now we will se what we can do with a rewound pzt strecther.
The pump was tuned back on afterwards.

Images attached to this report