Reports until 14:39, Saturday 01 March 2025
Heisenberg IFO Data Acquisition
hudsonl - posted 14:39, Saturday 01 March 2025 - last comment - 18:05, Sunday 09 March 2025(11905)
Excess Picoscope ADC Noise
Yesterday, the picoscope channel reading the breadboard OPO's DC differential channel contained a lot of excess ADC noise, and the recorder spectral shape of shot noise was not flat (see attached plot for Feb 28). This was resolved by closing the connection to the picoscope with the python command myPico_inst.close_scope(), unplugging the picoscope from power (a hard reboot), reconnecting the picoscope's power cable, and reconnecting to the picoscope with python. The same procedure should be followed in the future if the data recorded by the picoscope looks odd.

We also note that between the data taken on Feb 28th and March 1st, the BOPO homodyne's LO power level was increased from about 200 uW to 2 mW by swapping the outputs of a 90/10 beam splitter that were previously plugged into the wrong ports.
Images attached to this report
Comments related to this report
hudsonl - 18:05, Sunday 09 March 2025 (11909)
This can also be fixed by closing and re-opening the connection to the picoscope in python instead of unplugging it.