I tuned up the coupling from the 1% pickoff to the fiber for the squeeze angle control loop. Earlier this afternoon, the coupling was quite low (a few percent), so one of the optics on this path must have accidentally been moved. The coupling is currently greater than 90%, calculated as follows: We have 34 uW of IR power out of the CSQZ's OPO, so we should have 34 * 0.5 * 10^-2 uW = 0.17 uW of power in the 1% pickoff port (after the 50/50 BS and the 99/1 BS). The PD used to measure this power is a Thorlabs PDA20CS switchable gain InGaAs PD with a responsivity of 0.55 A/W and a transimpedance gain of 4.75 V/uW, so we expect a total response of 2.6 V/uW. The expected signal is then 0.17 uW * 2.6 V/uW = 0.44 V. We measure a signal of 0.40 V, indicating that we have 91% coupling efficiency.