Heisenberg IFO Data Acquisition
hudsonl - posted 14:29, Thursday 20 February 2025 (11893)
Heisenberg IFO Data Taking Procedure
We outline the data-taking procedure for the Heisenberg interferometer experiment. This post will be updated as the procedure is refined

(1) Align the setup with both squeezers in alignment mode. This is most easily done by running the 90% pickoff of the Lightwave through a 50/50 fiber beam splitter and sending one 50% path to the CLF port of each OPO.
     (a) Ensure that the voltages throughout the setup are set to the proper nominal values. The BOPO's OPO should be locked to a resonance with about 30 V applied to its PZT, the CSQZ's OPO should be locked to a resonance with about 125 V applied to its PZT, the squeeze angle control PZT should have an offset voltage of 75 V, and the MZI PZT should have an offset voltage of 125 V.
     (b) Align the beams out of the two OPOs to each other on the MZI's input 50/50 beam splitter
     (c) Align the MZI
     (d) Align beams to the CSQZ homodyne
     (e) Align beams to the BOPO homodyne
(2) Switch back to squeezing mode on both OPOs
(3) Close the PLL loop
(4) Take dark and shot noise references for each homodyne using the picoscope.
(5) Turn on each squeezer, set the OPO temperature to maximize NLG for a given pump power. Adjust pump power levels so the two squeezers have similar levels of NLG
(6) Close the CLF loop on each squeezer and the relative squeeze angle control loop.
(7) Lock the MZI by closing the MZI loop with the LB1005
(8) Lock the CSQZ homodyne to the CSQZ's CLF tone.
(9) Run an automated script to jointly optimize the values of the CSQZ homodyne demod phase angle and the relative squeeze angle control demod phase angle.
(10) Set the BOPO HD demod phase angle.
(11) Take data (with or without a modulation tone applied to the homodyne)