Displaying report 1-1 of 1.
Reports until 17:50, Sunday 19 January 2025
General
evanhall - posted 17:50, Sunday 19 January 2025 (11870)
Asymmetric Michelson for Teraxion testing
I set up a free-space Michelson with a one-way asymmetry of 10 cm and then aligned it with light collimated from the Teraxion.

The fringe contrast was 0.53 mW maximum and 0.14 mW minimum (58% visibility). This used a PDA05CF2.

With the interferometer near the half fringe, I drove the Teraxion's modulation with 0.1 V pp, which produced 0.104 mW pp. Using the relation Pmod / Pfringe = visibility * tau * 2 * pi * mod_depth (tau being the round-trip asymmetry 2L/c), this implies a modulation depth of 0.127 GHz pp, or a modulation coefficient 1.27 GHz/V. In comparison, the manufacturer's test report gives 0.99 GHz / V.

Changing L to 7.5 cm, the resulting power modulation was 0.09 mW pp. The fringes went from from 0.10 to 0.56 mW (70% visibility).

Changing L to 5 cm, the resulting power modulation was 0.06 mW pp. The fringes went from 0.09 to 0.57 mW (73% visibility).

Changing L to 2.5 cm, the resulting power modulation was 0.04 mW pp. The fringes went from 0.08 to 0.57 mW (75% visibility).

Changing L to 0 cm, the resulting power modulation was 0.01 mW pp. The fringes went from 0.00 to 0.63 mW (100% visibility).

Changing L to 12.5 cm, the resulting power modulation was 0.12 mW pp. The fringes went from from 0.15 to 0.49 mW (53% visibility).

A linear fit to these, correcting for the nonunity visibility and allowing for a free intercept to account for any systematic error in asymmetry, gives a coefficient 5.1 / m = 4 * pi * mod_depth / c, and hence mod_depth = 0.122 GHz pp, still with 0.1 V pp modulation depth, consistent with the above. The coefficient is therefore 1.2 GHz / V.

If either of the beams in the arms is blocked, no power modulation is visible on the scope.
Displaying report 1-1 of 1.