epolini - posted 08:38, Sunday 12 January 2025 - last comment - 11:38, Friday 17 January 2025(11858)
First Acoustic Noise measurements [Vienna]

[Chris, Eleonora]

Yesterday we installed a microphone (Razer Seiren v3 mini) inside the acoustic enclosure in the lab in Vienna (where the 50km interferometer is), between the two spools as shown in Fig. 1.

We acquired the noise floor spectrum turning off the microphone (using the Razer Synapse software) and the acoustic noise spectra with the AC on / off. 

Fig. 2 shows the data from the microphone with AC on and off in Volts.

Fig. 3 shows the data from the microphone with AC on and off in Volts/rtHz.

Fig. 4 compares the control signal from the loop (fast out) with the microphone data, with AC on and off. The microphone data are NOT calibrated, meaning that these data are useful only for a qualitative comparison of the peaks.

Images attached to this report
Comments related to this report
epolini - 13:19, Thursday 16 January 2025 (11867)

[Chris, Eleonora]

Today we also acquired the spectrum of the acoustic noise inside the box after turning on the compressor of the nanowire detectors. 

The attached figure shows the comparison of the past measurements with the one of today with AC on, compressor on and pump on.

 

Images attached to this comment
epolini - 11:38, Friday 17 January 2025 (11868)

I took also the measurement with the compressor on and the pump off. The structure at 1200Hz is gone as expected.

This is the acoustic noise spectrum when we want to measure single photons with the detectors in the current lab. 

I also investigated the shift discepacy between the curve taken at different times. I connected an other microphone and changed the volume level and it makes the level of the curve shift, as shown in the second plot attached. 

Images attached to this comment