Displaying report 1-1 of 1.
Reports until 17:37, Friday 03 January 2025
cSQZ General
hudsonl - posted 17:37, Friday 03 January 2025 (11850)
CSQZ Optical Path Stability
I put the compact squeezer in alignment mode after returning from winter break. After about two weeks with no activity in the lab and after building a box around the compact squeezer and likely shaking the device in the process, the visibility was still 93%.

After some re-alignment, I improved the visibility to 97.4%. 

Some observations from tuning up the alignment
(1) At high visibility, there's some degeneracy between the mirror's angle and polarization. It's important to jointly optimize these to achieve very high levels of visibility.
(2) The thorlabs tip-tilt mounts are much more stable when screwed into a post tapped for a 1/4-20 screw using a 1/4-20 screw as opposed to using a #8-32 screw with a washer and a post tapped for #8-32. I was unable to improve the visibility beyond 93% using the mount with a #8-32 screw + washer, and quickly achieved 97.5% visibility after switching to a 1/4-20 screw and mount.
Displaying report 1-1 of 1.