Notes for assembling picomotor rig

New Focus 8853 needs ball bearings for kinematic contact, in addition to custom mount.

New Focus 8742 needs new fuse (Littelfuse 229) and a new USB micro-B surface mount connector, which has been inexplicably ripped off the board (candidate replacement on Digikey: UJ2-MIBH-G-SMT-TR). Momentarily bypassing the fuse, the controller lights up but I did not attempt to drive it. The software installs and at least opens on Windows 10. We need a 3-to-1 picomotor concentrator cable (NF8725) unless we can find one around.

Comments related to this report
evanhall - 17:26, Monday 13 January 2025 (11863)

The kinematic mount fits fine into the picomotor portion of the mirror mount with 0.25″ steel balls.

The USB connection on the picomotor driver is not repairable and we'll have to use the ethernet port instead with a switch.