Reports until 16:40, Friday 13 December 2024
General
evanhall - posted 16:40, Friday 13 December 2024 - last comment - 13:31, Tuesday 17 December 2024(11839)
Teraxion laser initialization

Installed Teraxion LXM-S laser on breadboard in green lab. (LXM-S1550.00-040-03-I1-W0, serial number LXM533725, LIGO-S2401070)

Running the laser requires the installation of some software libraries and a National Instruments based control software with GUI. (This requred first installing the Microsoft Visual C++ Redistributable codebase in both x64 and x86.) The software runs on Windows 10 (maybe also 11, have not checked).

The measured output power is 42 mW.

The manufacturer confirmed there is internal isolation of at least 40 dB (this is probably what the test report means by return loss, which is measured at 60 dB). The interlock connector is MMCX series.

Comments related to this report
evanhall - 20:45, Monday 16 December 2024 (11840)

I measured the RIN by collimating the light with a 15 dB attenuator onto a PDA20CS.

It is probably better to remeasure with a custom photodiode, but that will require some retooling to get the correct transimpedance gain.

Non-image files attached to this comment
evanhall - 13:31, Tuesday 17 December 2024 (11841)

I measured this with a custom low-noise photodiode with a transimpedance of 2.4 kΩ. (https://dcc.ligo.org/LIGO-D2400261)

I am not sure why the RIN is 10× higher below 10 Hz compared to the PDA20CS measurement.

Non-image files attached to this comment