Facilities General
evanhall - posted 14:19, Thursday 12 December 2024 - last comment - 19:28, Friday 10 January 2025(11834)
Particle counts in NW22

Particle counts in three areas in NW22 are as follows, in units of 1/m3. Each is the average of three measurements using a Lighthouse Handheld 3013. This is with one person (me) standing about 2 m from the counter.

  Room CTN ADBC ADBC "night mode"
>0.3 µm 15 600 4300 6300 33 000
>0.5 µm 2900 890 970 6800
>5 µm 0 0 0 0

Can't really explain why ADBC night mode is so horrible. I think from the TUI documentation this is supposed to be a 25% reduction in the amount of blowing, by some measure.

We do not have readings at the >1.0 µm threshold, but the data we do have suggest the two clean rooms (on their "normal" setting, when the LEDs read 02 and ––) are better than class 100, worse than class 10.

Images attached to this report
Comments related to this report
evanhall - 18:06, Wednesday 08 January 2025 (11856)

Sławek, Evan

We measured the flow bench and some other areas with a different particle counter (Airy brand). Counts in 1/m3. The CTN clean room measurement was directly under one of the HEPA filters.

  Room Flow bench Room ADBC (off) CTN
>0.3 µm 9200 0 6400 13400 0
>1.0 µm 6700 0 1060 5300 0
>5 µm 3500 0 710 2500 0

 

evanhall - 19:28, Friday 10 January 2025 (11859)

I vacuumed in and around the flow bench's blower cabinet. The blowers are surrounded by porous black foam, which may be casting off black soot into the blowers. These should be replacted with proper pleated HVAC filters. We need two 20×24×1 filters and four 16×20×1 filters. Some effort is probably also needed to wipe out as much of the soot as possible.