Reports until 21:09, Wednesday 11 December 2024
Heisenberg IFO General
hudsonl - posted 21:09, Wednesday 11 December 2024 - last comment - 15:16, Thursday 12 December 2024(11832)
Heisenberg Interferometer Homodynes Broken?
As of yesterday (Dec 10th), the breadboard OPO's homodyne seemed not to work and had considerable excess dark noise. The board was using the M1 PCB initially. I swapped in the M2 and M4 boards, and neither of these behaved as expected either. However, the compact squeezer's homodyne board seemed to be working well and had about 30 dB of clearance between shot noise (with about 1 mW LO power) and dark noise. 

I left the electronics connected and turned on overnight and as of this evening when I came down to the lab, the compact squeezer's homodyne does not appear to be working either. There is only about 15 dB of clearance between shot noise and dark noise with the same optical power as before, and the spectral shape of the shot noise has changed (it is flatter out to higher frequencies). This leads me to believe that keeping the electronics turned on overnight caused the homodyne board to overheat, damaging the board. If this is the correct diagnosis, we should find a way to heat-sink the homodyne boards' voltage regulators without electrically connecting them to each other or to other grounds. 
Images attached to this report
Comments related to this report
hudsonl - 15:16, Thursday 12 December 2024 (11835)
[Hudson, Melissa]

Although the CSQZ homodyne appears to be at least partially broken, we measured its linearity to determine if it is at least usable for the time being. The attached plots show these measurements, which suggest its response is still reasonably linear.
Images attached to this comment