Reports until 22:13, Wednesday 27 November 2024
GRAVITES General
eoelker - posted 22:13, Wednesday 27 November 2024 (11829)
Prestabilization setup

[Eric, Dorotea]

Our new fiber feedthrough arrived this week, so we're finally able to recomission the laser prestabilization setup.  Below, some detail is included on the current status.  Note: so far, we've done no careful noise budgeting or characterization since this is only the 2nd time I've locked this loop.  However, I want to document everything in detail since we probably won't get back to this for a couple months.

The design for our prestabilization setup is based on this paper from ANU.  Since the NKT BASIK E15 has slightly worse free running low frequency noise than the laser used at ANU (Orbits Ethernal Slowlight), I've added feedback to the laser piezo to avoid range issues at low fourier frequencies.

The requirement for GRAVITES is to stabilize our laser frequency noise to below 2e4 Hz/rtHz at 10 mHz.  The performance attained at ANU is about 2 orders-of-magnitude better than what we require.  In principle, our design should be an improvement.  Our thermal time constants should be about a factor of 3 larger than reported by ANU, though we havent measured them directly yet.  We are also using a vibration-insensitive fiber spool which reduces phase noise from seismic vibrations by roughly 4 orders of magnitude according to our shaker tests.

Jaime's MSRP poster gives a nice overview of the setup.

 

Instructions for locking the servo

To lock the laser, the laser modulation must be enabled and set up with the proper parameters.  I attach a few screenshots showing the current settings: 1, 2, and 3.  Afer the laser is turned on, one needs to allow it to thermalize for about 15 minutes before trying to lock it.

The EOM modulation frequency has been set to 15.1 MHz, a frequency that gives decent sideband transmission for a 1km arm length imbalance when the carrier is locked to a dark fringe.  The demod phase was set to ensure that the servo has the overall correct sign.  We temperature stabilize the fiber EOM to limit RFAM issues.

The Marconi VCO is currently set to a frequency deviation of 20 kHz, DC coupled, the servo board has been set for maximum gain, and the laser piezo is currently being driven by a separate SR560, set as shown.  The Michelson is currently outputting ~50 uW of power onto an 1811.

Free running, we see the following on an oscilloscope.  Here, channel 1 (yellow) shows the DC power at the dark port, channel 2 (blue) shows the error signal after demodulation, and channel 3 (magenta) shows the correction signal being fed back to the PZT.  When the servo is unlocked, all 4 switches on the servo board are in the down position.  When you're ready to lock the servo, turn on the enable switch, then the two integrators.  When locked, the oscilloscope traces look as shown.  Ideally I'd like to see more suppression in the error signal to get a tighter lock, but im limiting the bandwidth to ~15 kHz because of the delay introduced by the 1 km arm length.  If I were doing things over again, I would probably lower the length imbalance by a factor of a few to increase the capture range and improve the bandwidth. 

 

Characterization of the servo (very limited so far)

Currently, I'm locking with a roughly 14 kHz bandwidth and 50 degrees of phase margin.  I'm being a bit conservative since the phase rolls off rapidly and I want to have enough gain margin.

The error signal spectrum may be bottoming out at a high white noise floor, though we need a noise budget to start making sense of things.  A back of the envelope callibration says this corresponds to about 2 Hz/rtHz.  The only loop component we've characterized is the servo input referred noise, which is a few orders of magnitude below this floor (see Jaime's Poster for a spectrum).  However we havent characterized any of the other components in the loop yet.  Everything else is second-hand and there could be some surprises waiting for us.  

Images attached to this report
Non-image files attached to this report