Eric, Wen
We've received all the hardware and mounted the 4-inch breadboard to level the beam height to the viewports (thanks to Kellen's help!) This setup avoids any periscope. We installed pump TRANS/REFL camera & PD, CLF REFL camera & PD, and seed TRANS camera & PD on the breadboard. With these sensors, we get the pump and seed pretty-well aligned into the VOPO cavity.
We then tried to lock the OPO loop. The demod PDH signal was too weak so we removed the 6dB attenuator of the 80-MHz injection to EOM. After that, we can see the signal pretty clearly on the scope. Locking with the current OPO servo board was not successful. We have to re-do some loopology as we did with SHG lock acquisition.