SQZ General
wenxuan - posted 16:14, Wednesday 16 October 2024 (11805)
Particle counts in LASTI SQZ cleanrooms

Eric, Wen

We did some housekeeping in preparation of putting the circular breadboard in the small cleanroom into the VOPO vacuum tank. We will continue our experimental work in the tank, which will take place in the large cleanroom so it better be as clean as the small cleanroom. We wiped all the surfaces that we could reach, including the curtains and all optical tables. The measured particle counts (0.3 micron and 0.5 micron) showed some improvements before/after cleaning.

 

Before cleaning:

Large cleanroom corners. South: 2/0; East: 26/2; West: 0/0; North: 5/0; Center: 8/2;

Small cleanroom: 2/0

 

After cleaning (we waited until the next day to let the dusts settle):

Large cleanroom corners. South: 2/0; East: 9/0; West: 0/0; North: 2/0; Center: 0/0;

Small cleanroom: 2/0