SQZ General
eoelker - posted 18:11, Saturday 12 October 2024 (11801)
Pump to OPO path solution

I measured the mode of the 532 nm collimator this afternoon.  This measurement was done using the old fiber feedthrough from the LASTI X HAM.  I made sure to coil the fiber a bit to ensure sufficient bending loss for any higher order mode content (these old homemade UHV fibers weren't quite single mode for green if memory serves).  I measured a waist disstance of .363 meters and a beam waist size of 156 um (radius).  Surprisingly, this gives > 99% mode matching to the VOPO if the fiber collimator is placed 12" away from M1 (VOPO input coupler).   Looks like we don't need any mode matching lenses on this path.  

For reference, here are the q values at the fiber collimator output:  q_x (used for mode matching) = -0.36342+0.14517i,   q_y =  -0.3708+0.158i