wenxuan - posted 18:58, Tuesday 01 October 2024 (11781)
Reviving LASTI SHG lock acquisition

Eric, Wen

 

We got pretty close to locking the SHG used for 16-m experiment. The SHG TEC was working, and the alignment was recovered from previous reports. We optimized the alignment again using the SHG TRANS while sweeping the PZT. 


Then we tried to lock the SHG manually with CMB. After a few unsuccesful attempts, we decided to take some transfer functions to characterize the SHG servo board. We identified each filter banks in the common path and slow path. The filters in the common path are working as described, except for the filter after the daughterboard in the common path that doesn't do anything. The filters in the slow path rail the error signals when engaged, which are consistent with my memories that we rarely use them at L1. Note that the overall CMB TF has a flat gain of 15 dB at low freq while the gain knob is set to the minimal of -32 dB. 

 

We then took a look at the TRANS PDH demod signal near the resonance. The SNR was not very good from the scope, so we increased modulation depth by adding another 13 dB of RF power to the SHG EOM. The cavity was pretty steady, and we took an OLTF of the plant while parking at the resonance (shown in attachment). The plant didn't have a 1/f roll-off as expected from PZT. The current CMB has a too high of a gain for us to set a UGF of a few hundred hertz. We decided to not decrease the RF power but change the filters of the CMB to be a better design, something like the site's CMB with a much lower pole freq. 

Images attached to this report