Mach-Zehnder Interferometer Alignment Procedure
Here we detail the alignment procedure for the Mach-Zehnder interferometer. This procedure was developed by Masaya during his time at MIT.

(1) Rough alignment: Roughly align the beams through the interferometer such that they intersect on the input and output beam splitters and have roughly the same polarization. In most instances, this will already be done and we can skip right to the precise alignment steps.
(2) Balance the input beam splitter: Remove the output beam splitter. Adjust the angle of the input beam splitter until the splitting ratio is exactly 50/50 by monitoring the power on the 1% pickoff paths on the differential receiver. This procedure is the same as aligning the beam splitter on one of the homodynes.
(3) Balance the output beam splitter: Use the arm on the upper left as the reference. Block the arm on the lower right using the flippable beam block. Adjust the angle of the output 50/50 beam splitter until it is exactly a 50/50 splitting ratio by monitoring the power on the differential receiver from the 1% pickoffs.
(4) Maximize visibility: Adjust only the lower right interferometer arm to maximize the interferometer's visibility. Using a pair of mirrors, we can adjust the position and angle of this beam to match it to the upper right reference arm. 

Note: While adjusting the beam splitters' splitting ratio, it is essential to simultaneously maximize the power on the diodes using the steering mirrors to ensure that the beam is centered on the diodes.

After these steps, the interferometer's visibility should be >99%.
Comments related to this report
hudsonl - 11:47, Tuesday 01 October 2024 (11780)
We have observed that the interferometer's alignment is fairly sensitive to the exact resonance location we're locked on as the MZI is aligned. As a result, we should warm up the OPO to the phase matching temperature (nominally 34 degrees celsius) and lock on a consistent co-resonance location for both squeezing and aligning the interferometer.