[Wen, Eric]
Today we took the old FC16 6" dichroic viewport from HAM-X and relocated it to the VOPO tank in preparation for building the new 10dB squeezing setup in the highbay.
The 6" viewport was originally attached to a 10" to 6" reducer flange. Since I couldn't find a cleaned/baked 6" blank we decided to remove the reducer flange as well. We installed a 10" blank to close things up again
We applied 190 in-lb torque to both VOPO and HAMX flanges, according to T1100066