jacques - posted 22:37, Thursday 19 September 2024 (11764)
Swapped custom RFPD with Thorlabs PD on GR PDH

It was previously noticed that the SNR on the BSQZ's PDH signal was quite low (0.01Vrms of noise against comparable Vpk of PDH error signal). To increase the SNR, I swapped the RFPD with a Thorlabs PDA10A2. This significantly improved the responsivity (V/W) therefore the quality of the error signal, so much that I had to go from an ND1 to an ND3 filter before the PD, and reduced the "quadrature_factor" on the iq module from 50 to 1. Here is a screenshot of the PDH sweep now, the TF and the spectrum (UGF of 1.5kHz).

I also did some quality of life improvements on the PDH plots, we can now choose which resonance to plot and the voltage units both reflect the RP's output and the actual PZT voltage. I will push the modifications in the file "lockBreadboardOPO.py"