Anthony Amorosi - posted 10:57, Thursday 19 September 2024 (11761)
HAMX-ISI control update

This takes over the updating of the HAMX-ISI control system done by Morgane Zeoli from Aug. 5 to Aug 23 after the CDS upgrade. 


•    Adding DQs channels:
T240 and ground STS2 sensors signals have been added in the model as DQ channels. The procedure is as follow:

1/ Kill the model by ssh-ing to m1hamx and using the following commands, in this order, 
rtcds stop m1isihamx    →   rtcds stop m1hpihamx   →  rtcds stop m1iophamx                
(killing the iop may not be necessary)

2/ From a machine with matlab 2019, enter the name of the channel to add as DQs in the comment window in the simulink .mdl file, followed with its sampling rate. Update the .mdl file.
Example: SENSCOR_GND_STS_X_MATCH_IN1_DQ 2048

3/ Restart the model with the following commands, in this order, 
rtcds start m1iophamx    →   rtcds start m1hpihamx   →  rtcds start m1isihamx                

4/ Reboot the front end by ssh-ing to m1fb0 and using the following command,
sudo reboot 

The STS2 signals can be accessed at the following name
M1:ISI-HAMX_SENSCOR_GND_STS_{X,Y,Z}_MATCH_IN1_DQ              , it include calibration
The T240 signals can be accessed at the following name
M1:HPI-HAMX_SENSCOR_{X,Y,Z}_IN1_DQ                                                 , it include calibration


•    Gains
Gains on some sensors have been noticed wrong. Some indications were:
- The unity gain frequency being 10-ish Hz instead of 30 Hz, filters have not being changed.
- Signals amplitude between GS13 and CPS after blending were not matching.

Transfer function from ISO_X_Exc in ISO_X_IN1 was showing a DC offset by a factor of 1/2.27. Assuming the CPS and actuators are calibrated, it has been decided to apply a gain of 2.27 on the signal of the GS13. This gain is added as “Gain_corr” in the M1:ISI_HAMX_GS13INF_{H1-3,V1-3} filterbank.

Accrodingly, the STS2 sensor on the ground for sensor correction also has a 17.93 gain correction switch in the  M1:ISI-HAMZ_SENS_COR_GND_STS_{X,Y,Z}_IIRHP filterbank under the name “Gain”.

Accordingly, the T240 semsor on the ISI table also has a 0.432 gain correction switch in the M1:HPI-HAMX_STSINF_B_{X,Y,Z} filterbank under the name “Gain_corr”.

With all these gains ON, the control are stable and with the expected unity gain frequency. The origin of these gains is still unknown and I would recommend carefully identifying the acquisition gain and ensuring that these are correct.

 
•    Tilt-coupling compensation
Morgane previously identified the coefficients that minimized the  X to RY coupling. These were implemented in the CPS_ALIGN matrix as X to {RX,RY} elements. These has been switched for {RX,RY} to X, as it makes a little more sense: i.e. X motion sensed by the CPS is sent to the X actuotor, bu also in RX and RY, instead of sending the the X actuator motion sensed in RX and RY. 

The values have been fined tuned, currently sitting at X to RX = -0.004 and X to RY = 0.000. 
Values updated in the /home/controls/Data/Populate_Hamx_ISI_MEDM_Screen.m script.

Y to RX and Y to RY have not yet been investigated.

•    Sensor correction:
 Sensor correction has been implemented and optimized to a certain level. 
The gain leading to the best performance was found to be 1.125. It has been implemented as the “Boost” control switch in the m1:ISI-HAMX_SENSCOR_GND_STS_{X,Y,Z}_MATCH filterbank.

Details of the measurement cam be found in the diaggui 24_08_30_Senscor.xml file in /home/controls/Data/2024_09_18_Anthony/, where the sensors signals has been recorded stored for different sensor correction gains.

References are as follow:
  0 to 11:  control off
12 to 23: control on, no sensor correction
24 to 35: sensor correction gain =  1.0
36 to 47:                                     =  1.5
48 to 59:                                     =  0.5
60 to 71:                                     = -1.0
71 to 83:                                     =  1.25
84 to 95:                                     =  1.125
currrent:                                     =  1.075

/!\ This was done with the 250 mHz blend. The ISI is running now with the 100 mHz blend, I am unsure of if the blend frequency could impact the optimization of the sensor correction gain. The ISO controllers in X,Y,Z have also been upgraded after this test.

•    ISO controllers:
It has been seen that the ISO filters (Controllers and boost) could use some improvement in X,Y,Z. New versions have been produced and are stored in the “Cont_4” and “Boost_4” switches in M1:ISI-HAMX_ISO_{X,Y,Z}.

Related to the gain issue mentioned above, the controller gains had to be tuned so the unity gain frequency measured on diaggui was matching the designed value of 30 Hz. The correction is directly implemented in the “Cont_4” switches, and the value was different for the three designs.

The measurements that served the design of the controllers can be found running the Load_OL_TF.m matlab script in /home/controls/Data/2024_09_18_Anthony.

An attempt of a v4 version of the controller for RY has also been produced, but the closed-loop system rings at a frequency of 40-ish Hz probably due to a wrong notch in the controller desgin. This version is aborted for now.

•    Feedforward
Transfer function from feedforward sensors to (ST 0 L4Cs) to the platform have been identified. 

The set of measurement can be found in /home/controls/Data/2024_09_18_Anthony/24_09_09_FF_TFID.xml
It contains the set of transfer function from the M1:ISI-HAMX_FF_{X,Y,Z}_EXC drive to M1:ISI-HAMX_BLND_GS13{X,Y,Z}_IN1 and from M1:ISI-HAMX_FF_{X,Y,Z}_IN1 to M1:ISI-HAMX_BLND_GS13{X,Y,Z}_IN1 with no drive.

The feedforward transfer function is obtained by taking the ration of these two quantities. The /home/controls/Data/2024_09_18_Anthony/FeedForward_FilterDesign.m matlab script loads the data and attempt a fit on the FF transfer function.

The Y and Z filters have been implemented in CEDM in the M1:ISI-HAMX_FF_{X,Y,Z} filterbank under the “Cont_1” control switch, performance are not satisfying as the phase of the fitted filter only matches the experimetnal phase in a limited frequency range from 5 to 10 Hz.

The main assumption that may explain the poor performance I that the HPI stage is not operating (not damped), this causes a strong resonance peak at 10 Hz making the fitting of the FF transfer function difficult. We might consider locking the HPI stage for now and retake the measurement.


•    Set of ISI ASDs and TF
A set of reference ASDs have been taken with the current design of the control filters. The measurements can be found in /home/controls/Data/2024_09_18_Anthony/24_09_09_RefASD.xml

The references are as follow:
00 to 02: Control OFF
03 to 05: controller version 3, sensor correction, 250 mHz blend.
15 to 17: controller version 4, no sensor correction, 100 mHz blend.
18 to 20: controller version 4, +sensor correction, 100 mHz blend.

The /home/controls/Data/2024_09_18_Anthony/ISI_Perf.m matlab script loads data of the ISI with different isolation levels and plots spectrum and transmissible transfer functions with the nost recent designs of the controllers/blends/gains.

The /home/controls/Data/2024_09_18_Anthony/ISI_Spectrums.m matlab script allows to fetch data at a specied GPS time and plot ASDs.


•    Vacuum
The chilling system for the vacuum pumps has been repaired and the pumps are now operating normally. The rough and turbo pumped has been turned on few times for a couple of hours in the week Sept 09 – Sept 15 and 1e-2 Torr vacuum has been reached. No strong evidence of leak was observed.

During the first pump down of the system however, it was observed that the ISI control loop went unstable at a frequency of 68 Hz. It was suspected that the removal of air made a rocking mode very high Q, which is driving the ISI into instability. The door has been opened and the T240 vacuum pod, which was initially clamped on the ISI, has been placed on 3 thin shims to make it kinematic. This assumption turned out right as the 68 Hz mode disappeared after closing the tank and pulling vacuum again, and the control loop were stable.

•    Next steps
For the next steps, I would suggest:
- closing investigating the cause of the gains mismatch, and ensuring that the corrections that were applied were correct;
- Have a second look into the tilt-compensation matrix, identify the couplings in Y direction;
- Redesign the version 4 of the RX, RY and RZ controllers;
- Lock down the HPI stage and redesign the feedforward filters;
- Have the Turn_on. M matlab script working.
 

Images attached to this report