Heisenberg IFO General
jacques - posted 00:04, Thursday 19 September 2024 (11760)
Breadboard SQZ's LO locked using the pair of AOM instead of the PZT

Jacques, Hudson, Melissa

We previously had issues on the CLF-LO loop because we were running out of range with the PZT (3um of actuation range), due to the natural drifts of the optical paths between SQZ and LO. Furthermore the PZT's actuation motion was affecting the alignment of the LO. To mitigate the range of the drifts, Hudson removed the green fiber but it did not seem to significantly improve the range issue.

We then used a pair of fiber AOMs (from the old cSQZ "fiber box") in place of the PZT to close this loop. We swapped the 90/10 BS for CLF/LO, with a 50/50 BS. We have to reoptimize the CLF's incident polarization, as well as the LO's as the fiber AOMs have an efficiency which is polarization dependent. We used the moglabs's channels 1 and 2 to send two signals at 150MHz to upshift and downshift the LO, with actuation (freq mod) on ch 1. This means that the cSQZ no longer has a RF source to drive its AOMs.

Today, I sucessfully locked the CLF-LO loop of the BSQZ using the setup consisting of two AOMs on the LO path and actuation on the frequency modulation. For the non-CLF related locks, I used the jupyter notebook "BSQZ general script". I used the jupyter notebook "debug AOM lock" to close the CLF loops, this is a copy of "clf2.py" with minor modifications. After optimization it will be copy pasted back into clf2.py. See picture attached for the TF and noise spectrum of this new loop. The main change was to increase the gain to ~500.

Using this I was able to get 4.5-5dB of SQZ vs 13 dB of aSQZ. I verified that the NLG temp was optimal and realigned the HD2 visibility to be 96.8%. So I don't really understand why we can't get more squeezing right now.

The next goal is to fine-tune this new loop and do a CLF calibration sweep on the BSQZ and compare it to the model. It would also be good to calibrate the noise spectrum in phase noise units.