I have now connected both HDs to the acquisition RPs and have taken shot noise vs electronic noise spectra. They are taken with the usual 2.6mW of LO on HD1 and 1.1mW of LO on HD2. I have properly isolated the HD board using baluns and 50 Ohm adapters.
They are here.
Conclusion: on both HDs, there is about 20dB (21dB for HD1, 19 dB for HD2) of clearance below 1MHz. The electronic noise has some periodic pulses. It's not clear whether they come from the environment (we have already put all the electric plugs onto the same set of wall plugs, but it is not isolated from the rest of the grid), or from the setup itself.
I used the jupyter notebook "CLF calibration acquisition" to acquire the data and "CLF calibration data/CLF calibration data analysis" to analyze and plot the data.