hudsonl - posted 16:49, Wednesday 11 September 2024 (11754)
Tune Up Breadboard OPO CLF Path
We noticed that several of the optics on the path from the IR fiber coupler to the breadboard OPO were loose. We noticed this after one of these optics was bumped and the path was misaligned. I went and tightened all of the optics on this path, re-aligned the beam through the AOMs, and re-aligned the beam into the OPO. 

To align the beam to the OPO, I sent green light through the OPO's input/output coupler. I then aligned the CLF beam reflected off of the CLF's input mirror to the green beam. This gave me clear cavity flashes on a camera and clear transmission peaks on a transmission PD. From there, it was straightforward to fine-tune the alignment to the cavity. Currently, the mode matching is nearly 100% for the 532 resonances corresponding to the lowest piezo drive voltages (the mode matching degrades for larger PZT voltages as the mirror is not actuated exactly parallel to the cavity axis).

For the AOMs, we have 12 mW of incident IR power before the first AOM and about 6.3 mW in the first diffracted mode after the second AOM, so the efficiency is about 72% per AOM.
Displaying report 1-1 of 1.