I've uploaded the plots and data of the HD boards M1, M2, M3 taken with the Agilent in the dropbox. Data is missing on M1 RF2 (I had changed a component) yet I had checked directly on the Agilent that this data was similar to the other boards. As for board J, a lot needs to be resoldiered and data should be retaken since I haven’t followed every change and I guess some components differences can be the cause of a sharper drop in transfer function. As for M4, it consumes 0.15A instead of 0.09A; I haven't tried to debug it even roughly but this might be a quicker way to have another board than building another one from scratch. As for the data files and their titles: -they refer to RF output ports when "RF" is written; to DC OUT (main) port when not. -some values of incident light power is written for the transfer functions yet the spot was larger than the photodiodes so we should approximately divide this power by a factor 10^3 -the transfer functions were done with one of the PD inputs only, cf a previous logbook post on the symmetry in the PD inputs of these boards