Displaying report 1-1 of 1.
Reports until 13:34, Sunday 25 August 2024
cSQZ General
hudsonl - posted 13:34, Sunday 25 August 2024 (11741)
CSQZ Switch Between Alignment and Squeezing Modes
Here are the steps to switch the compact squueezer between alignment and squeezing modes. The whole process should take about 30 minutes and only requires swapping two fibers, which should both now be easily accessible.

To go from "squeezing" to "alignment":
(1) Swap the 90% output from the fiber bench with the EOM for the SHG to the input of the polarization paddles on the CLF path. Turn off the AOMs on the CLF path and cover the output of the second AOM with a fiber cap.
(2) Check the DC power on the OPO reflection RFPD for the CLF 2f signal. The PD saturates at 10 V and ND filter should be added as needed to prevent saturation.
(3) Connect OUT 2 of the RP 28 red pitaya to the J8 pin on the analog 10x amplifier
(4) Send the output of the CLF RFPD through a DC block and then to the input of RP 28 instead of to the bias tee
(5) Replace the 60k resistor in series with the OPO mirror's PZT with a barrel

To swap from "alignment" to "squeezing":
(1) Replace the barrel in series with the OPO mirror's PZT with a 60k resistor
(2) Send the output of the CLF RFPD to the bias tee instead of through the DC block and the input of RP 28
(3) Connect the output of the OPO control red pitaya to the J8 pin on the analog 10x amplifier
(4) Swap the 90% output from the polarization paddles on the CLF fiber path to the fiber bench with the EOM for the SHG. Clean the fiber at the output of the AOMs and connect this to the polarization paddles.
(5) Remove any unneeded attenuators from the CLF 2f RFPD. Use the polarization paddles to maximize CLF power, then adjust the RF level out of the Moglabs RF driver to attenuate the CLF power to about the right level and then use the polarization paddles to fine-tune the incident CLF power. Note: this could instead be done with a VOA if we want to add one.


Displaying report 1-1 of 1.