Reports until 11:49, Thursday 22 August 2024
Heisenberg IFO General
hudsonl - posted 11:49, Thursday 22 August 2024 (11733)
CSQZ Switched Back to Squeezing Mode
[Hudson, Jacques]

We switched the compact squeezer back to "squeezing mode" with high power on the SHG pump path and lower power on the CLF path. The power levels are currently as follows

SHG Pump: 75 mW out of fiber --> 64 mW incident on SHG after losses from PBS and 90/10 BS (note: there is 125 mW of opwer after the Faraday but only about 60% of this couples into the fiber connected to the free-space coupler. These fibers could probably be spliced to reduce losses here).

CLF port: 1.2 mW out of fiber with both AOMs driven at 100 MHz, 32 dBm

CSQZ Homodyne LO: Up to 5.2 mW out of fiber (can attenuate this with a VOA)

Squeeze angle control LO: up to 3.7 mW (can attenuate with a VOA)

See Masaya's post here for the full fiber schematic: https://emvogil-3.mit.edu/lasti/index.php?callRep=11722

While making the swap, Jacques and I also rearranged the physical location of some fibers to try to make the fiber path more clear by eye and keep all of the PLL fibers and signals in the same place. The monitor for the PLL was moved to the same scope as the other PLL signals and the high bandwidth PD used to monitor this signal was moved to the PLL optics board on top of the diabolo.