masaya - posted 19:15, Monday 19 August 2024 (11731)
Beam size of 2nd Homodyne is put back and small enough
[Jacques, Masaya]

As Jacques reported, the lenses to make the beam small at the photodiodes of 2nd Homodyne detector are not put on the correct location when we measured 6 dB squeezing.

Now, the situation is currently back and two beam sizes on the photodiodes are the same as before (180-190 um).

The attached pdf reads that the beam sizes were 160 um and 200 um diameter when the 6 dB squeezing was obtained.
The attached photos show the beam sizes and the optical setup for the measurement.

We put the Homodyne board and aligned the light to be incident on it. We successfully confirmed the intensity plateau.