jacques - posted 03:07, Monday 19 August 2024 (11730)
Second HD shot noise limited but lens issue resulting in no alignement plateau

After debugging the HD board on Masaya's side by using a correct pair of PD, I tried to check that it could be shot noise limited by sending some power on the LO side and looking at the spectrum of the DC out once the light is aligned on the two PDs.

I obtain the photo attached, on the Agilent. Red is the Agilent noise floor, White is the HD dark noise, Yellow is the shot noise spectrum with 0.333mW of LO power and no squeezing/antisqueezing. The plot goes up to 10MHz. This shows that the HD is indeed shot noise limited.

It is important to note that I was not able to find an intensity plateau using the lenses that Masaya had installed just before the HD. This means that the beams are not fully inside the active area of the PDs, thus they are clipping. This is consistent with a beam profile measurement where I measure a beam waist diameter of 600um (Gaussian fit) at the level of the HD PDs, which only have an active diameter of 500um.

Therefore one should change the lenses preceding the HD, so that the waist diameter is at most ~200um, to minimize clipping on the HD PDs and SQZ loss.