Reports until 22:02, Sunday 18 August 2024
Heisenberg IFO General
jacques - posted 22:02, Sunday 18 August 2024 (11728)
Debugged fast homodynes

Malo, Jacques

Yesterday evening we debugged the fast homodyne as Hudson, Eric and Masaya had trouble understanding why one of the PD paths was never responding to any incident light.

It turned out that the high quantum efficiency PDs which were used for the tests have their anode and cathode's pins positions different from the previously used high quantum efficiency PDs and the PCB board layout. This is easily tested simply by taking one of the PDs, hooking two of the three PD pins to a scope, send some light on the PD and see for which combination of pins the voltage read on the scope is positive (it's easy to get several 100s of mV using a simple lightbulb as the light source). 

We therefore swapped the pair of PDs which had a different pin layout, with another pair of high quantum efficiency PDs (same 500um active diamter and footprint as the original PDs on which the HD worked fine). Doing this allowed to solve all issues related to these HD and they are now fully functionnal again.

Conclusion: There are two batches of high QE PDs, which look exactly the same but have different pins layouts. In the future, before using an unknown PD, one should always verify that its pin layout is compatible with the layout of the corresponding electronics board. When in doubt, do the simple test of sending light on the PD and making sure which pin corresponds to which theoretical port of the PD.