Reports until 19:20, Thursday 15 August 2024
Heisenberg IFO General
hudsonl - posted 19:20, Thursday 15 August 2024 (11726)
PLL Power Supplies
[Jacques, Malo, Masaya, Hudson]

We were unable to close the PLL loop earlier this evening. After some fiddling around with the electronics, we realized this was due to the PLL relying on three different power supplies

(1) The +- 24 V power supply by the PLL setup powers mode of the PLL electronics
(2) The +18 V power suppl by the PLL setup powers the PLL's high voltage amplifier
(3) The compact squeezer's electronics box powers a 23 dB signal amplifier on the RF path from the RF PD to the PLL's input

The big takeaway is to make sure the compact squeezer's electronics box is powered in addition to the PLL's dedicated power supplies when trying to engage the PLL.