Displaying report 1-1 of 1.
Reports until 19:31, Wednesday 14 August 2024
Heisenberg IFO General
masaya - posted 19:31, Wednesday 14 August 2024 (11725)
Beatnote signal of CLF from Breadboard OPO and LO beam for Squeezing Angle Control
Successfully obtained the beatnote signal for Squeezing Angle Control.

The CLF transimission power is 24 uW and LO power is 0.46 mW when the measurement was done.
The obtained peak at 3.1 MHz has -39 dBm and the noise floor is -82 dBm.
Images attached to this report
Non-image files attached to this report
Displaying report 1-1 of 1.