[Vicky, Eric]
We've been making steady progress the last several days getting the LASTI squeezer back online. Today we tackled the Beckhoff. Initially we were unable to get data from any of the beckhoff hardware channels. SQZ 1, 2, and 3 were all failing to initialize properly in the TwinCAT system manager (were staying in INIT NO_COMMS state instead of going to OP state). We were able to track this down to a couple of broken EtherCAT modules in the SQZ 1 and SQZ 2 chassis. We replaced the broken module in SQZ 1 but will need to order additional spares to fix SQZ 2.
We disabled the middle rail in SQZ 2 for now and bypassed the broken EtherCAT module in the chassis. Once we did this, all of the remaining rails are now functioning properly and we are seeing the corresponding hardware channels in EPICS. The middle rail in SQZ 2 controls the OPO common mode board, so we'll have to wait on recomissioning the OPO servo until I'm able to replace the broken EtherCAT module.
I placed an order for some spare modules on amazon, they should arrive in a couple of weeks