Reports until 23:53, Monday 05 August 2024
cSQZ General
jacques - posted 23:53, Monday 05 August 2024 (11719)
PLL not locking on the correct side: sign issue on the slow loop

Malo had previously noticed that the PLL, when locked, corresponded to the "wrong side" of the lockpoint (beatnote of 2*51MHz = 102MHz between the two lasers instead of locking them at exactly the same frequency). I made a fiber setup to directly probe what the beatnote with the two lasers is on a 150MHz Thorlabs PD sent to the Agilent. I solved the issue of the lock setpoint by changing the sign of the slow loop. As of this evening, the command that locks the PLL on the correct side is:

PID_response_full(offset_slow=0.112, prop_fast_final = 0.018, integral_slow =  0.001, integral_fast_final = 35, swap_gain_signs=True, ramp_up_gains=True)