I spliced together the interferometer for the GRAVITES laser frequency prestabilization chamber. I'm getting around 97% visibility in the Michelson after balancing powers with the VOAs. Should be sufficient based on the reference we're aiming for. I also wired up the temperature sensors for both shields and the heater for the outer shield baseplate. Just a few tasks left before we can close up and pump down:
To confirm the suspicion of a vacuum chamber leak, I verified that the pressure no longer decreases when the pump is activated but rises progressively over time.
At the beginning of last week we measured 1.9 torr while at the end of last week 2.95 torr and today 3.95 torr.