[Eric, Eleonora, Jaime] Today, we continued working on the Michelson interferometer for the pre-stabilization setup. First, we viewed the fringes on the oscilloscope and adjusted the VOAs in each arm to maximize visibility, which is currently at 79%. We then reduced the AOM drive from 24 dBm to 5 dBm to avoid saturating the signal mixer. Additionally, we incorporated the temperature controller for the EOM to stabilize the phase modulation. Next, we added an inline polarizer after the EOM, which allowed us to properly measure the residual amplitude modulation. While monitoring the demodulated error signal(image attached), Eric noticed an oscillation with a frequency of about 1 Hz, corresponding to a frequency drift in the laser of 200 kHz/s. This drift lies outside the VCO’s range of 100 kHz. Finally, I conducted a power budget for the setup while driving the AOM at 5 dBm: After first Faraday isolator-67.6mW After 90/10 beam splitter to AOM-4.43mW After AOM-79.0μW After 50/50 beam splitter to second Faraday isolator-26.90μW After second Faraday isolator-24.84μW After EOM-13.10μW After Polarizer-11.44μW At photodiode-≈2.3μW Tomorrow, we will tune the integrators on the servo board I made and act on the PZT of the laser to correct the drift.