jacques - posted 02:04, Sunday 14 July 2024 - last comment - 10:40, Monday 15 July 2024(11703)
Up to 7dB SQZ and 13dB aSQZ on cSQZ

Malo, Jacques

The goal of today (Saturday July 13th) was to finish the cSQZ recovery and measure SQZ. We started from yesterday by putting the moglabs' RF outputs at 105MHz sharp and 95MHz sharp.

The first task was to debug the CLF and LO loops. Indeed, with the new fiber disposition and HD board, the error signal amplification/ attenuation had to be rethought for these two loops. I added a function plotErrorSignal() in the clf.py script. This allows one to get a plot of any input demodulated. Doing so on the CLF-pump error signal ("in2" on redpitaya27 with demod_freq =  20e6 Hz (see below for this small offset)) and the CLF-LO error signal ("in1" with demod_freq = 10e6), I realized that both were saturating because they were above 1V in amplitude. We therefore had to remove an amplifier on the CLF-pump cable path, and add a -20dB attenuation on the CLF-LO cable path (out of output "RF2" of the new HD). Then, we can see that the demodulated error signals are spanning +- 0.6V, which means that they are correctly calibrated.

The interaction between the OPO's IR coresonance plot and the CLF REFL PD goes as follows:

1/ reduce the CLF power using the paddles so as to see a nonsaturating IR resonance when the OPO is scanned (use opo.measureIrResonance() to get the IR plot only)

2/ tune the tempearture of the OPO while scanning to have the transmitted CLF almost overlap with the green resonance (today it was at 32°C).

3/ lock the OPO

4/ Increase the CLF power again to max, or such that it is visible on the transmission PD, simultaneously, optimize the CLF parametric amplification using tec.opoSetpoint(31.95). This is sensitive to the 0.01°C level.

5/ Close the CLF-pump loop.

6/ Then, if there is too much CLF preventing one from measuring a correct SQZ level, reduce the CLF power by using the VOA (do not use the paddles; they should always be in the optimal polarization position in order to reduce the polarization fluctuations).

7/ Reclose the CLF-pump loop if necessary.

In the meantime, Malo optimized the visibility on the PD by putting the CLF at zero detuning when the OPO was locked on coresonance and out of parametric amplification, reaching about 96% visibility on the HD. See the three scope pics here, here and here for the measured max/min voltages and the bias voltage of the HD output. We then tried to close the CLF and LO loops with similar PI gains than before (demod frequencies respectlively 20MHz and 10MHz) and look on the spectrum of the DC port of the HD to see if there was any squeezing. While we indeed saw that the spectrum dipped below shot noise, it was clear that the noise level was slowly periodically oscillating between SQZ and aSQZ, on the scale of several tens of seconds. This was strange since we thought that all the loops were closed properly: I could plot the demodulated error signals of the CLF-pump and CLF-LO using "plotErrorSignal(...)" and I indeed saw that they were around 0 when the loops were closed. 

We then decided to quantify this periodicity by plugging the HD DC on the scope and measuring any periodic signal over the course of several minutes, with all the loops closed. In a perfect world, since there is always the beatnote between the LO and the 10MHz CLF, we would expect a constant amplitude, 10MHz oscillation. This was not what we observed, as seen on the picture of the scope attached here. Rather, we saw a very periodic oscillation of the CLF-LO beatnote, with frequency ~25mHz. 

I looked at the LO PZT voltage shown on the thorlabs driver and saw that it was stable over this oscillation. This lead us to believe that this slow oscillation comes from the CLF-pump loop, which effectively was making the SQZ/aSQZ quadratures rotate at this slow frequency. 

We tried to investigate this effect further, by changing the CLF-pump loop's demodulation frequency by 25mHz, i.e going from "demod_freq = 20e6" to "demod_freq = 20e6+0.025". This is further motivated by the facts that

1/ the CLF-pump loop's modulation and demodulation devices are separate and not clocked together: the modulation is made by the moglabs, while the demodulation is made by the RP.

2/ we know that the moglab's displayed output RF frequency is not reliable, as it had been noticed several times before that the moglab could produce a signal that is offset from the displayed frequency for no apparent reason.

Doing so, and locking the CLF-LO loop without any offset from 10MHz, allowed us to stabilize the SQZ spectra. By lowering the CLF power (we did not try to reduce it as much as possible), we were able to resolve between 6.5 and 7 dB of SQZ, versus up to 13.5 dB of aSQZ. In absolute values, we had -118dBVrms/rt(Hz) of shot noise, <-150dBVrms/rt(Hz) of electronic noise, -104.5 dBVrms/rt(Hz) of aSQZ and -124.5 dBVrms/rt(Hz) of SQZ. See attached plots here and here. While the aSQZ level is comparable to the max which was previously measured on the setup (makes sense since the GR power is more or less the same), the SQZ level is significantly better than before, which implies that the HD mode-matching was the limiting loss contribution for SQZ observation.

I will push the code I used today on the git.

TBD1: change the opo.lockGRonly function for the user to choose their resonance. In the future, only use this function to lock the OPO in order to bypass the annoying CLF power management to lock the OPO on coresonance.

TBD2: push the CLF power even lower (possible since with max CLF power we have no problem detecting the CLF-pump and CLF-LO error signals) to see if it further lowers the observable SQZ floor.

TBD3: optimize the loops params, reduce the PDHs' EOM amplitude modulations to reduce the overall phase noise contamination.

TBD4: pump power scan to obtain a full loss model of the cSQZ. 

Images attached to this report
Non-image files attached to this report
Comments related to this report
jacques - 10:40, Monday 15 July 2024 (11704)

On the matter of the REFL CLF PD and the coresonance plot: In the future we should use the TRANS OPO IR PD for the IR coresonance plot, instead of the REFL CLF PD. This is because

1/ We already use TRANS OPO IR PD to optimize the OPO temperature to maximize the parametric amlification when locked

2/ The CLF is entering through the undercoupled port of the OPO so resolving the resonance on REFL means distinguishing a ~5mV peak among a ~1V signal, which either makes the RP saturate if the off-resonance part is too strong, or prevents it from seeing any resonance feature because it's too weak. Both of these issues are solved by the TRANS PD because:

a) it is in transmission so the off-resonance voltage is ~0V (i.e. no risk of saturating the RP)

b) it has an internal tunable gain so the on-resonance voltage can be adjusted, usually we can see a peak of ~50mV (i.e. no risk of not seing anything on the RP)