We continued the recovery of the cSQZ which started yesterday. We managed to go up to the CLF-Pump and LO-CLF locks, effectively seeing averaged antisqueezing on the HD.
I changed the code of the opo.lockGRonly in order for it to work. It now can lock on GR error function only without needing a coresonance of the IR. I noticed that the opo.measureIr and opo.measureErrorSignal functions did not stop the scan at the end of their excecution, which made stand-alone error signal visualization inconsistent between two consecutive executions of the functions.
It was suspected that the issues related to the CLF REFL PD not consistently outputing an IR resonance on the opo.lock() is due to filtering of the IR resonance scope signal, through the clean_data function. I added parameters which allows one to adjust the lowpass and highpass frequencies in order not to filter the resonance peak. Furthremore, I filtered the first 5% of the data in order not for the lowpass to make artifact peaks appear at the begining of the plot.
Because the CLF REFL PD is looking at the undercoupled port, we have that this resonance signal will not go above ~6mV, out of a ~60mV out of resonance signal (on the scope it's 60mV but there is a x10 amplification before the RP input). Furthermore one needs to be careful that the DC power on this PD never exceeds 100mV on the scope, so that the RP input does not saturate.
To conclude on te IR coresonance signal, it seems that there is no straightforeward way to increase it for free at the moment. We can nevertheless proceed with the current setup.
On the matter of the OPO, the demod phase of the opo.lock seems to be crucial for the stability. However, the opo stability was overall fragile today; I could not find a setpoint such that I could tap on the OPO without breaking the lock, unlike the previous times. Because the error signal seems to sensitive to the demod phase, I suspect that the main green peak that we see when the opo is scanned might not actually only correspond to the 00 mode. Another possibility is that the SHG's 50MHz sideband is also present on the GR. This should be easily alleviated by reducing the modulation amplitude on the SHG's EOM.
I put the CLF at 10MHz detuned from the carrier. This means that there is more loss from the AOMs, such that about 0.5uW of CLF is transmitted out of the OPO on coresonance. Nevertheless, we are able to see it on the RF2 port of the HD.
At some point the new HD board that we were using seemed to have broken, since the optimally balanced situation had a DC out fluctuate by several 100mV. We did not inverstigate this HD further, we simply swapped the electronics board with the other new HD that we had built; we kept the same PDs. Now, with this new HD board, the signals are consistent again: the DC can go to 0, we have <-150dBVrms/rt(Hz) of electronics noise vs -118dbVrms/rt(Hz) of shot noise with 2mW of incident LO power.
Since the BS had to be tuned after this swap, the CLF path is misaligned with the LO. This should be tuned further by unlocking the CLF and maximizing the RF2 beatnote at 10MHz, or look at another signal at DC and put the CLF's detuning close to DC.
To get the CLF-pump phase 20MHz error signal, we had to add a minicircuits amplifier after a bias tee (to get the RF output only).
To get the CLF-LO 10MHz RF2 error signal, it seems that we do not need any amplification, provided that LO and CLF are already well-aligned.
We then tried to close the CLF-pump and CLF-LO loops. Transfer functions could be taken and they made sense, but we could clearly see on the RF2 signal that the amplitude of the (supposedly stabilized) signal is changing over the course of several seconds. This means that these loops are not actually closed in a satisfying manner.
We left the system with the CLF at 0 detuning.
TBD: add a timeout on the clf.lockCLF() in order not to wait for the input acquisition if it's too low in power (we had -60dbm of 20MHz beatnote without amplification)
TBD: add a script on the CLF 20MHz loop to plot an error signal scope trace when the loop is open vs when it is closed.