GRAVITES Optics 1550nm
epolini - posted 18:29, Thursday 11 July 2024 (11696)
Measurement of axial and radial acceleration of the vibration insensitive spool - 1st attempt

We built a setup similar to the one in the COL paper (doi: 10.3788/COL201917.081403) to measure the acceleration of the new 1km spool. We did the measurement using a balanced MZI with DC readout, using the new homodyne. We added the 1km plastic spool, a VOA on one arm and the AOM on the arm that was unbalanced before. We set the AOM power to -22dBm to have 100nW from both arms recombining at the homodyne inputs.

We placed the new spool inside the box I build one month ago on a table I took with Rich from Myron's office. Rich, Jaimie and I installed two shakers on this bench. We resumed the power supply (most of the boards inside are damaged and does not work, so please use only the one already connected that we verified works properly). Fig. 1 shows the shakers, the guralp and the spool on the new table. The power supply is under the table. 

Measurement calibration

The calibration in terms of 1/g is given by:

mag_cal = (mag_lin) *   (vco * guralp * rad_cal) * 9.81 / (2 * np.pi * freq**2 )

where mag_lin is the magnitude in linear scale, vco is the calibration of the vco from V to Hz (10kHz freq dev - 6.5kHz/V), guralp = 1375 V/m/s, rad_cal =  (2*np.pi) / (n * lam * L) with lam = 1550e-9, n=1.45, L=1km.

The bench was shaking in all the directions also optimizing the positioning of the shakers, the guralp and the spool. After some attempts to get a TF between the correction signal of the loop and the guralp signal (both HOR and VER), we decided to substitute the table with a stiffer one and do the measurments with an UMZI configuration.