Reports until 00:14, Wednesday 26 June 2024
cSQZ General
jacques - posted 00:14, Wednesday 26 June 2024 - last comment - 14:26, Thursday 27 June 2024(11686)
Attempt at dither lock the OPO: why does the REFL error signal have a static offset?

Following a discussion at the group meeting yesterday, it was suggested to simply dither lock the OPO, since we are only interested in seeing a beam out of the cavity, and not interested in any noise contamination about squeezing; although we may want to be somewhat careful about possible differential amplitude noise in the CLF vs LO; although^2 I would not be so afraid of this on the visbility since the dither is fast compared to the fringes.

Thus the idea of this loop is to send a 80kHz dither on the NPRO's PZT, demodulate either on TRANS or on REFL of the OPO (as seen from the CLF's path), and actuate on the OPO cavity mirror's PZT. Utilmately, one would like to lock on REFL since the lock on TRANS would happen along the alignment process to the HD (we don't have any pickoff since this is also the SQZ path).

One can actually theoretically show that in dither mode, the error signal slope should be identical whether one looks at TRANS or REFL. In the linear regime it is proportional to the excursion in frequency of the dither and the input power.

I did the necessary cabling to make this dither loop on redpitaya28.mit.edu. In particular:

1/I had to unplug the +10x amplification of redpitaya24.mit.edu's output and plug in 28's output in order to have the correct sign sent to the PZT Driver.

2/I had to add a -10x amplification for the dither drive sent to the laser PZT. This is ok since the max signal sent to the laser is +-10V at 80kHz, whereas the manual recommends not going above +-50V at 100kHz. 

I first attempted to lock on TRANS. This was successful with the following parameters of IQ module (see screenshot). See screenshot for the error signal as seen in TRANS (the TRANS PD is put on +30dB internal amplification).

I then tried to look at REFL. The difficulty of this is that much more DC power is dumped into that PD (5mW before filter vs 50uW on TRANS), which means that the error signal needs to be extracted with sufficient amplification but without saturating the 1V input of the RP. I get the following error signal (see screenshots for the plots over the whole PZT range and zoomed and without the DC offset). Although one can clearly see an error signal, because of the DC offset I can't lock onto it in good faith. Tweaking the code to take the offset into account should allow me to lock but this is a very concerning behavior of the demodulated error signal. Why is there a DC offset on the demodulated error signal? 

Conclusion at this stage:

1/ We have enough modulation depth to see an error signal by laser dither, both in REFL and TRANS.

2/ TRANS locks but REFL is DC-offsetted and thus can't lock. 

Next things to try: 

- Switch the RP input to high voltage setting and remove a ND filter in the REFL PD path.

- Find a BNC Tee or lowpass to kill the DC signal of the REFL PD and thus increase SNR on the error signal.

- Lower the modulation amplitude of the laser PZT to see if the DC offset comes from some kind of spurious amplitude modulation. Likewise, change the modulation freq.

- Tweak the code to lock regardless of the DC offset?

 

Images attached to this report
Comments related to this report
jacques - 14:26, Thursday 27 June 2024 (11689)

For reference the manuals of the PD that are used in REFL (PDA10CF, no switchable gain) and in TRANS (PDA100A, with switchable gain).

Non-image files attached to this comment