We want to PDH lock the OPO in reflection via the CLF port. To do so, we want to use a high modulation frequency, since the OPO's IR bandwidth is about 50 MHz. On the PDH locks for the SHG and OPO in normal operation, we get a reasonable PDH error signal by driving a broadband EOM directly and at frequencies less than 60 MHz. Admittedly, these EOMs should be made resonant as well. We were unable to get good SNR on the OPO's PDH lock via the CLF port using this technique. To improve the lock's SNR, we used the LIGO EOM impedance matching board https://dcc.ligo.org/LIGO-D1200794 We shorted L4, so the impedance in the LC resonator was 270 nH. We used a Newfocus 4004 EOM with a nominal capacitance of 20 pF, giving a predicted resonance frequency of about 68.5 MHz. We measured a resonance frequency at about 99.5 MHz, indicating a true capacitance value of about 10 pF, roughly consistent with the manufacturer's specifications. The circuit has a monitor port, which can be used to take its transfer function. An image of this is attached. The Transfer function from the input to monitor ports has a gain of about -8 dB at resonance. As measured by Rich Abbott, the ratio between the output and monitored voltages should be about 130:1 or 42 dB, indicating that the circuit's gain at resonance is about 34 dB, or that Vout = 50 x Vin.
I assume that the OPO's mirrors have 1064nm power reflectivities of R10 = 0.89 and R20 = 0.9995, while they have 532nm power reflectivities of R1GR = 0.96 and R2GR = 0.9999 (cf table 4.1 in Alvaro's thesis). I neglect the cavity losses. From Black's equation 3.3, the complex error signal is epsilon = r[omega]r[omega+Omega]* - r[omega]*r[omega-Omega]. The demodulated error signal is maximized at a demodulation phase, which yields a maximum equal to |epsilon|.
The difference between the SQZ mode and the HD alignment mode is that the input and output mirrors are effectively exchanged, since the GR and CLF enter through the opposite sides of the cavity. While for GR the input was R1GR and output was R2GR, for CLF the input is R20 and output is R10. Thus, I find a reduction of error signal (assuming same input powers and modulation depths) of a factor of 1/637.
If I take into account that it has been measured that the GR finesse is actually comparable to the IR one (about 50), this means that T1GR (proportional to the GR finesse in the approximation of large finesse) increases by a factor of 3, which means that R1GR is comparable to R10. Then I get an error signal attenuation of a factor of 1/213.
This means that, to retrieve the same error signal voltage, ceteris paribus sic stantibus, one would need to increase the modulation depth by a factor of 213 from what was sent to the GR EOM. In practice we should already get an appreciable SNR on the demodulated error signal if we increase the modulation depth by a factor lower than that.
See notebook screenshot for computations.