cSQZ General
hudsonl - posted 11:10, Friday 14 June 2024 (11674)
Mode Matching CSQZ Seed Laser and PDH Locking OPO in IR
[Hudson, Malo]

In normal operation, when producing squeezed light, we send most of the light to the input of the SHG and a small amount of light through the back of the OPO's high reflector for the CLF beam. However, it is useful to send a large amount of power through the CLF port of the OPO for alignment. Previously, we had kept the OPO locked during alignment by locking the SHG with a small amount of IR light, locking the OPO with an even smaller amount of green light, and then keeping the OPO dual-resonant by adjusting the phase-matching temperature. Instead of this scheme, we would like to lock the OPO in IR directly, which should considerably improve the lock stability and the power level out of the OPO.

To do this, we added a free space EOM phase modulator in the path of the main laser, which will be used for the IR PDH lock to the OPO. We also added a free-space Faraday Isolator, which will allow us to remove a fiber Faraday, and re-tuned the mode matching. Previously, we had about 50% mode matching from the main laser into the first fiber. After adding the new optics, we have improved this to about 70%. Now, with 2.9 mW out of the main laser (after an attenuator formed by a half waveplate and a polarizing beam splitter) we have about 2.4 mW sent to the fiber (after some losses from the Faraday, EOM, waveplates, lenses, and mirrors) and about 1.45 mW out of the 90% pickoff from the 90/10 beam splitter the laser is fiber-coupled to. Accounting for the 12% loss of the 90/10 from the perspective of the 90% port (10% from the pickoff, 2% additional loss), the fiber coupling is 1.45/(0.88*2.4) = 70%.

Currently, about 50% of the light from the main laser is coupled out of the 90% port of the 90/10 beam splitter, so when the laser power is turned up to 500 mW, we expect 250 mW out of this port. After the fiber bench, we then expect 125 mW of light. Currently, we have a Thorlabs Faraday circulator with about 3dB loss after the fiber bench, so we expect about 60 mW pumping the SHG. After replacing this with a higher quality AFW circulator, we expect about 100 mW of light pumping the SHG. The AFW Faraday is specc'ed at 0.8 dB expected loss and we expect about another 0.2 dB from the fiber connector (this could be eliminated by splicing).

We expect the OPO to reach threshold at about 80 mW of incident IR pump power (see https://emvogil-3.mit.edu/lasti/index.php?callRep=11660 ), so we should be able to laze the OPO with the new configuration once the new Faraday from AFW arrives.