Hudson borrowed a Faraday circulator from Slawek for some quick test; to see if adding a circulator between the output of the UBench and the pigtail fiber reduced the power fluctuations, as it seems that most of the back-reflected light couples back to the UBench. Sure enough, as I had written in my previous entry, this qualitatively solves the issue: when there is 0.92mW incident on the cSQZ, we measure 0.17mW of light at the unwanted output of the circulator. This is compatible with the figure of 10% of relative power fluctuations that I was observing previously on PD13, and this shows that such amount of light does indeed reflect back into the fiber. For example, now, tapping on the waveplates on the UBench does no longer induce any significant power fluctuation on PD13.
The bad news is that this Thorlabs circulator has intrinsic losses of 2dB, which means that, with the usual 50% mode-matching of the UBench, we have 5.4 dB of losses between the input of the UBench and the input of the cSQZ. But we can still lock the SHG and OPO in HD alignment mode, provided that the laser power is maximized.
Next step: buy a fiber FI (ideallly with as low losses as possible) and make sure to tape the fibers of the FI on the table to minimize power fluctuations due to polarization changes in the fiber.