jacques - posted 23:12, Tuesday 04 June 2024 - last comment - 13:58, Wednesday 05 June 2024(11665)
Towards NLG scan (1)

I mainly spent the day trying to understand Hudson's code, change it slightly for ease of access, and recover all the cSQZ loops to prepare for a NLG scan tomorrow. The weather today in the lab was rather stable. I borrowed the Govee thermometer/hygrometer from Eleonora's setup and placed it on the cSQZ. It registered 21.2+-0.2 °C and 52+- 1% throughout the afternoon.

I modified the opo.lock() code to quadruple the scan speed (decimation of 256 instead of 1024 on measureIrResonance and measureErrorSignal), to symmetrize the error signal sweep (see screenshot) and have a better slope fit.

I have noticed that sometimes the opo.lock() does not acquire the IR trans signal, this is because the CLF power is too high. Turning it down solves the problems and allows the lock to proceed.

I then measured a NLG with the GR power that we are currently able to produce (25mW out of the SHG; if this value goes down, check the IR alignment on the SHG by tuning the input PBS). I was able to measure an amplification factor of 17 using the power on the PD blocking the HD path (amplified on resonance= 2.62V, non amplified on resonance = 0.32V, non amplified out of resonance = 0.176). This was obtained by tuning the OPO temperature to tec.opoSetpoint(33.07) when the OPO is on resonance.

I was not able to lock the clf loop at first, because the gain was too high. I proceeded to change the code of clf.lockClf() so that it can accept two arguments (proportional gain and integrator), without having to reexectue the whole code. For example, with clf.lockClf(prop=7, integral = 5e2), I get the following transfer function using clf.clfNA() when the loop is closed (see screenshot). Actually I don't really have another way to check that the loop is closed other than verifying that the TF makes sense.

Similarly, I was not able to close the LO loop at first, because the gain was too high. I added the gain and integrator params to the lock function. For example, with clf.lockLo(demod_phase = -10, prop = -10, integral = -8e2), I get the following TF using clf.loNA() (see screenshot). I can also check that the loop is not saturating by looking at the RP correction to the piezo and making sure that the PZT driver output does not saturate at ~107V.

I then look at the SR to see where the asqz/sqz is. To change the LO phase, I tune the demod_phase of clf.locLo(...). By carfully maximizing the OPO temperature and the LO phase, I managed to get up to 14 dB of asqz (see picture attached), but the SQZ level was still at about 2.5dB below shot noise (not shown here but an analysis will follow). There are two conclusions from this and a thing to test:

1/ The system seems to be loss dominated

2/ The current power budget seems enough to generate squeezing levels for our upcoming experiments. This does not mean that we should be satisfied with the power fluctuations.

3/ I did not try to lower the CLF power further but it could be that this is again a limiting factor on the amount of SQZ.

Next steps are to check 1 and 3 by optimizing the HD visibility and lower the CLF as much as possible, since so far all the loops involving the CLF seem fine.

Images attached to this report
Comments related to this report
jacques - 13:58, Wednesday 05 June 2024 (11666)

HD sqz/asqz plot from the data obtained yesterday. There is no peaks in the region 50 to 70 kHz, from which we can assess 14.05 dB aSQZ and 2.65dB SQZ.